Calibration

Calibration Procedure for Electronics on Muon Lifetime

In this section we set up the SRS DG645 Digital Delay Generator for testing the efficiency of our digitizer. The DG645 may be thought of as a function generator which is specifically designed to generate very precise, synchronized pulses. You should first take a look at the manual of this device and briefly familiarize yourself with its capabilities. Quantitatively, to what precision is the timing of the pulses guaranteed by the manufacturer? Is it good enough for our purposes? The manual for this generator may be found online.

Setup the DG645 as follows:

Set the Delay Generator to trigger internally at 25kHz (every 40 $\mu s$). If the Muon Program Range is set to something other than 40$\mu s$, this rate should be changed to match it. To do this:

Set the generator to create a 100ns pulse 1 $ \mu s $ after the internal trigger fires. To do this:

Now set the offset of the AB pulse to 0 and the amplitude to 0.5V. To do this:

The AB output of the Delay Generator should now be outputting a 100ns pulse every 40 $ \mu s$. Plug the digitizer card into the AB output of the Delay Generator. Open Scope Soft Front Panel, and view the signal. If the signal matches what you expect, you have set up the Generator correctly.

IMPORTANT NOTE: the output of the Delay Generator expects a 50$\Omega $ load. If the load impedance is much larger, the signal amplitude will double. The digitizer card may be configured to have a 50$\Omega $ input impedance or a 1M$\Omega $ impedance. If you are using Scope Soft, make sure the input impedance is 50$\Omega $ (enable DC coupling, click Channel and select the input impedance to be 50 $\Omega $). When using a separate Oscilloscope, use a T junction with a 50$\Omega $ terminator. The Muon software (View Signals and Muon Detection) both set the impedance to 50$\Omega $ automatically.