Experimental Procedures

Contents

Microscope How-To

BMC_Station319

The BMC Station

Setup2

Block Diagram

You will use the Zeiss Axiovert 200 inverted compound microscope illustrated above. Starting from the top right of the diagram, the microscope has a variable intensity halogen light source. A condenser focuses light from the halogen bulb onto the sample under observation. A rotating turret in the condenser holds several irises and stops that are used for different contrast methods such as bright- and dark-field. (As shown, the condenser iris is selected, as would be the case for normal bright-field illumination.)

The objective nosepiece rotates to select one of four objective lenses: 5x, 10x, 20x, and 40x. Although they are depicted as simple lenses in the diagram, the objectives are actually sophisticated, compound optical elements. More information about objectives can be found in references 6 and 7. Specifications of the objectives may be found at the Zeiss website.

Light enters the microscope body through the selected objective and is optionally split into two paths. A setting wheel selects whether 0, 50, or 100 percent of the light is diverted to the left sideport for the camera. The remaining light goes on toward the eyepieces. A CCD camera is mounted with its sensor at the focal plane located just outside the sideport. Images captured by the camera are transferred to the PC via a Firewire connection.

A binocular tube splits the image headed toward the eyepieces in two for direct observation with both eyes. The eyepieces are 10x, providing overall magnification of 50x, 100x, 200x, or 400x depending on which objective is selected.

Look over the detailed schematic of the Axiovert 200 microscope. You will notice some differences between the microscope in the drawing and the one in the lab – most significantly in the condenser (described below). To become familiar with the microscope controls, follow the instructions below to make a slide with a combination of 10µm and 0.44µm polystyrene spheres and then view it in both transmitted light and dark-field illumination. But first some general warnings:

* The microscope is a delicate instrument. It is never necessary to force any adjustment.
* Do not remove any parts from the microscope (except for eyepieces) or attempt to clean the optics. 
* If the microscope needs cleaning, ask a GSI or Don for help.
* Do not use canned air on the microscope. It can leave a difficult-to-remove residue. 
* To lengthen the lifetime of the halogen lamp, turn off the light source with the HAL on/off switch (Schematic-7) 
  when it is not in use.

How to pipet

The pipettes are located against the wall in a special holder (suspended above the shelf, about eye level)

Forward Technique: Fill a clean reagent reservoir with the liquid to be dispensed.

  1. Depress the push button to the first stop.
  2. Dip the tip under the surface of the liquid in the reservoir to a depth of about 1 cm and slowly release the push button. Withdraw the tip from the liquid touching it against the edge of the reservoir to remove excess liquid.
  3. Deliver the liquid by gently depressing the push button to the first stop. After waiting about one second to fully empty the pipette tip, continue to depress the push button all the way to the second stop. This action will empty any remaining liquid.
  4. Release the push button to the ready position.

If necessary, change the tip and continue pipetting.

Never allow liquid to enter the body of the micropipette

Make a viscous suspension

A solution will consist of three parts, water, which makes up most of the solution, a solute (either PVP or glycerol) to provide the viscosity, and the beads which we will then observe.

Water can be located in the large squeeze bottle marked “Deionized Water,” be sure to use this rather than tap water as tap water will fill your slide with various unwanted particulate matter.

The Solutes can be found in the upper cabinet and are also clearly marked. Use the long and thin metal scoops/spatulas to measure out the PVP and the large-volume pipette to measure out the glycerol. Note that the glycerol is extremely sticky (ie viscous) and will not be measured precisely using the pipette as it will stick to both the inside and outside of the filter tip. To measure both the PVP and glycerol it is best to use the scale located at the lab station.

Computing viscosity

Use this table to compute the required dilution of the solvent. The PVP in the lab is in pure powder form. The stock Glycerol is a thick liquid with >99% purity. It will require some care to measure the pure glycerol accurately, since it tends to stick to the sides of the pipette tip. For this reason, weighing the glycerol before adding the water is probably a better technique than adding the glycerol to the water.

PVP and Glycerol Viscosity versus Concentration

Viscosity (cP)      PVP to Water Ratio (by weight)         Glycerol to Water Ratio (by weight)
1.66                0.50% : 99.5%                          18.0% : 82.0%
2.50                1.00% : 99.0%                      32.2% : 67.8%
4.65                2.00% : 98.0%                      44.2% : 55.8%
13.2                4.00% : 96.0%                      64.5% : 35.5%

You may either use these values or interpolate between them (are the relationships linear?).

Creating Polystyrene Suspensions

Beads are located (and should be kept) in the refrigerator. Each of the vials is clearly marked with the size of bead that it contains. (Note: The sizes reported on the vials are mean particle diameter, not radius.) These vials are often extremely concentrated and you may wish to create your own diluted solution to work with.

Note that the densities vary to a small extent between the different size particles. For this reason, the following procedure may need to be adjusted slightly for each of the different particle sizes.

Repeat this process with each experimental condition selected.

Make a slide

Now that we have a proper vial of viscous bead solution made up we need to transfer a sample of it onto a slide so that we can observe the beads’ behavior.

Microscope Figures

BIO_image14

Place hand on position number (1) then gently tilt back top of microscope

BIO_image13

Putting a Slide on the Microscope Stage

BIO_image15

Place microscope slide between the arrows shown above

BIO_image16

Slide installed onto microscope stage

Microscope

Schematic of Axiovert 200 Microscope

View the slide in transmitted light

Microscope images look best when properly illuminated. The procedure developed by August Köhler’s is nearly universally used to achieve uniform illumination with little reflection or glare and minimal sample heating. In this step, you will view your sample under Köhler illumination.

As shown in Figure 4, there are two irises in the illumination path of the microscope. The one on top (closest to the light source) is called the field iris. The condenser iris is a little farther down. The adjustments for both irises are highlighted in Figure 2. The condenser iris adjustment is located at the edge of the turret disk near the bottom of the condenser. You will adjust both of these irises to achieve the best image.

Condenser_Tilt_Back_0771

Figure 4: Picture of Field Iris, Condenser Turret, and Condenser Iris Adjustment

Fig. 2 is a picture of the actual condenser assembly you will find in the lab. The markings on this condenser are a bit confusing since it was designed to be used mainly in upright microscopes instead of inverted models such as the Axiovert 200.

It is necessary to go through this adjustment procedure every time you change objectives or samples.

* It is possible to make the tip of the objective collide with the slide. 
* If you crash an objective into the slide, you will very likely break the slide. 
This will release the liquid from the slide on to the objective and make it very dirty. 
Try very hard not to do this. 
* When you get close to the proper focus, it is an excellent idea to engage the focus limiter 
at a setting slightly higher to reduce the chance of running in to a slide.
* Avoid focusing on slower particles because they are close to the edges.

Adjusting the field and condenser irises

Iris

Fig. 5. Adjusting the irises

Do not skip this step. The microscope must be focused on a sample to properly set up Köhler illumination.

Why do the PS spheres appear bright in the middle?

Set-up darkfield illumination

Darkfield

Fig. 6. Principle of dark-field illumination. all the illuminating light is rejected, and light scattered by the sample is collected by the objective

Dark field illumination allows for observation of small particles under the microscope that would not normally be visible. This is due to the scattering of light onto the particles. The effect is similar to looking at stars during the daytime versus night.

Test_0_frame_18

Figure 7: Darkfield image of single 100 nm diameter gold nanoparticles

Viewing / Tracking Particles in Viscous Suspensions

Now that suspensions have been made for each experimental condition, and the microscope has been fully configured, we are now ready to take data.

Estimating the Pixel to micron conversion factor

Prepare a solution of 10 µm PS beads in the same manner as the viscous suspensions were prepared, except use deionized water instead of viscous solution.

You should obtain values of around 0.3-0.5 µm/pixel for the 20x objective. You will use your value as the conversion factor in the software.

Obtaining Data from viscous solutions

Repeat this for each experimental condition selected.

Note: Watch out for bulk flow. If you see a number of particles moving in one direction, they are likely undergoing bulk flow. This could be due to evaporation of liquid from beneath the coverslip or an air bubble popping, or various other conditions. If you see bulk flow occurring, your data will be skewed. Make sure there is no bulk flow when collecting data.

This should conclude Part I of the experiment

Tracking Particles in a Living Cell

Making an onion slide

Onioncellpicture1

Objective 3 (40X magnification) image of an onion cell. The red square is enlarged in the next picture. The onion's cells possess an interesting property which is that the particles tend to use the cell walls like highways, and tend to travel along these walls (although they will not solely be traveling down the cell walls

Onioncellpicture3

Objective 2 dark-field image of the onion cell. The intracellular particles are now white dots

Onioncellpicture2

Bright-field image magnified image. The red circles with arrows indicate the particles that are traveling down the cell wall

Onioncellpicture4

Dark-field zoomed-in image of the onion cell. The red circles with arrows indicate the particles (now appearing as bright white dots) that are traveling down the cell wall.

Making Observations

First you should spend a little while looking around, trying to find some regions of interest. Start with bright-field illumination (Objective 3, upside-down 2 to the left of the condenser). Adjust the brightness of the lamp as needed, along with the focus. Note the different types of movement and where they tend to occur. In particular, be sure to investigate regions around the cell walls, regions around the nucleus, and also see if you can find anything happening within the otherwise empty center of the cell. Most of the activity happens on the lower and upper layers of the cell as the center is occupied with the vacuole which should be devoid of anything except water. If you scan through the depths of a few cells carefully however (using the focus knob to move in depth) you should be able to find isolated actin fibers which make for very clean data-taking. Next, try dark-field illumination (Objective 2), and compare your camera images to the pictures above for both illumination regimes. Remember that each time you change the objective (and slide, for that matter), you must refocus the objective using the large knobs on the side.

Your data analysis will be much easier if you can isolate the forms of movement within the cell and only take data on one type at a time. Otherwise you may have to manually sort through particle trajectories.

Take data using the particle tracking software (Brownian Motion Application) as you did with the viscous suspensions.

This should conclude Part II of the experiment